| Model Number |
FP7 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) for the reported lot was also reviewed, and no issues were found; the product was manufactured, tested, and released according to validated procedures.As the device has not yet been returned, no device malfunction could be confirmed.
|
| |
|
Event or Problem Description
|
|
Not satisfying result in iol reduction.
|
| |
|
Additional Manufacturer Narrative
|
|
The explanted valve underwent visual inspection and functional testing.No issues were found with the device during visual inspection.A steady state pressure test was also conducted on the unit to mimic the physiological flow conditions of the human eye, where the resulting pressure in the system measured and yielded passing results.The returned valve met the acceptance criteria and performed as expected.The issue of high iop as reported by customer could not be replicated or confirmed.
|
| |
|
Search Alerts/Recalls
|