Model Number FRX521H14C
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
A dreamstation 2 cpap device was returned to a third-party service center in reference to a complaint of using up the water.During the initial evaluation of the device, the third-party service center visually inspected the unit and found the pca had thermal damage.It was also noted that the blower was defective, and the pollen filter needs to be replaced.There was no allegation of patient harm or injury.The device was repaired by the third-party service center.The pca, blower, and pollen filter were replaced.The device was serviced, inspected, and tested by the third-party service center and passed the testing requirements set by the manufacturer.The manufacturer is submitting a final report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key22550350
Report Number2518422-2025-107978
Device Sequence Number14458375
Product Code BZD
UDI-Device Identifier00606959062757
UDI-Public(01)00606959062757
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K200480
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberFRX521H14C
Device Catalogue NumberFRX521H14C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2024
Initial Report FDA Received Date07/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-