• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. SENSATION 7FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. SENSATION 7FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0469-01
Medical Device Problem Code Difficult to Flush (1251)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
Customer called stating he is new to iabp therapy and caring for its patients.He states he cannot flush the inner lumen of a fo iab.He states he has talked to the physician, and they are unsure on how to proceed.The therapy is running as expected on a cs300 pump.Patient was involved.No harm.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id (b)(4).
 
Additional Manufacturer Narrative
Updated fields - b4, d9, g3, g6, h2, h3, h6 (type of investigation, investigation findings, component code, investigation conclusions), h11.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION 7FR. 34CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key22552220
Report Number2248146-2025-0000394
Device Sequence Number3903927
Product Code DSP
UDI-Device Identifier10607567106755
UDI-Public(01)10607567106755(10)3000351710(17)261117(11)231117
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0469-01
Device Catalogue Number0684-00-0469-01
Device Lot Number3000351710
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date07/20/2025
Supplement Report FDA Received Date08/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CS300.
Patient SexUnknown
-
-