| Model Number |
366131 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.We can confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The passeo-18 peripheral balloon catheter was selected for treatment of a mildly calcified lesion (70 % stenosis degree) in a mildly tortuous part of the proximal sfa.The affected device was introduced into the patient's body and positioned accordingly but the balloon could not be inflated.After device withdrawal, a leakage was observed.Another balloon was used to successfully finish the procedure.
|
| |
|