• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. SMARTPATH TO DSTREAM FOR XR AND 3.0T; FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. SMARTPATH TO DSTREAM FOR XR AND 3.0T; FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET Back to Search Results
Model Number SMARTPATH TO DSTREAM FOR XR AND 3.0T
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a report that the quadrature body coil (qbc) seal was damaged.This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the magnet covers.If this seal is damaged it is likely that this could lead to serious injury.
 
Additional Manufacturer Narrative
The qbc seal that was damaged was replaced and the system was handed back to the customer for use.The failure of the qbc seal to remain in place is considered a malfunction.Remedial action: a field action was initiated under 2023-pd-mr-013.A field safety notification was sent to the customers informing them about this potential issue.The field correction has since been made available to the field in dec 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTPATH TO DSTREAM FOR XR AND 3.0T
Common Device Name
FULL-BODY MRI SYSTEM, SUPERCONDUCTING MAGNET
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs street
cambridge, MA 02141
MDR Report Key22553075
Report Number3003768277-2025-006593
Device Sequence Number9647469
Product Code LNH
UDI-Device Identifier00884838108684
UDI-Public(01)00884838108684
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K043147
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSMARTPATH TO DSTREAM FOR XR AND 3.0T
Device Catalogue Number782145
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2025
Initial Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2023-PD-MR-013
Patient Sequence Number1
Patient SexUnknown
-
-