| Catalog Number |
ECS29B |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent 7/21/2025 d4 batch # unk b3: only event year known: 2025.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure that product description is 29 mm ethicon circular stapler xl sealed.The stapler did not fire.There is no adverse impact patient/user was noted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent 8/8/2025 d4 batch #197d53 corrected data d4: udi#.Investigation summary the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample revealed that the ecs29b device arrived with the anvil missing.The breakaway washer was not present and there were no staples present.The device was reloaded with staples; a new washer was placed on the device and it was tested for functionality with a test anvil.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Although it could not be determine the cause of the reported incident, proper usage of the proximate ils circular staplers is important to ensure functionality.For more information, please refer to the instructions for use.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 197d53, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|