| Catalog Number |
TSVM6B8 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
|
| Date of Event |
04/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number: (b)(4).A4: patient weight unknown / not provided g4: additional pma/510(k) number k101880.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that implant was removed due to bone loss and infection.Previous rct-tooth, which had large periapical infection and subsequent sinus inflammation extracted and grafted.Implant surgery completed with concurrent grafting.Underwent healing.At time of uncover, buccal dehiscence noted with very little reverse torque, implant was explanted uneventfully.Patient will be opting for traditional fixed denture.Tooth site: 3.Symptoms as a result of the event: inflammation.
|
| |
|
Search Alerts/Recalls
|
|
|