CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS SYS W/CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX824T
Medical Device Problem Codes Infusion or Flow Problem (2964); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 06/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an m.Blue plus (#fx824t) was implanted during a procedure performed on (b)(6) 2024.According to the complainant, the shunt system caused an overdrainage and had adjustment difficulties.The patient underwent a revision procedure on (b)(6) 2025.No patient complications were reported as a result of the revision procedure.The complained device was returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damages of the valves were determined.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves do not operate within the permissible tolerance in their respective relevant position.An accelerated outflow of progav 2.0 and a slower outflow of the m.Blue could be determined.Adjustment test: the valves were tested and the m.Blue is adjustable, but the progav 2.0 is not adjustable to all pressure settings.Braking force and brake function test: the brake functionality test has shown that the brake function of the m.Blue operates as expected and the braking force is within the given tolerance.However, the brake function of the progav 2.0 operates as expected, but the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, organic deposits were found.Results: according to the results of the investigation, a reduced outflow of the m.Blue was detected.Furthermore we found the progav 2.0 was no longer adjustable and had an accelerated outflow.The visible deposits in the shuntsystem have led to the functional deviations.We can determine visible deposits in the control reservoir.The deposits had no effect upon the specifications at the time of the examination.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.The examination of the return has been completed with the drafting of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS SYS W/CONTROL RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key22554184
Report Number3004721439-2025-00204
Device Sequence Number9559437
Product Code JXG
UDI-Device Identifier04041906504830
UDI-Public4041906504830
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFX824T
Device Catalogue NumberFX824T
Device Lot Number20071556
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Initial Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient SexFemale
Patient Weight90 KG
-
-