Model Number FX643T
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 06/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damage of the valve were determined.Permeability test: a permeability test has shown that all components are permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 operates within the accepted tolerance in the horizontal position.The shuntassistant 2.0 does not operate within the specified tolerances in the vertical position.An accelerated outflow of shuntassistant 2.0 could be determined.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, bloody deposits were found in progav 2.0.Results: based on our investigation results, we can determine an accelerated outflow at the shuntassistant 2.0.No deposits were visible in the visible area.However, when opening the shuntassistant, it is only possible to see part of the interior.There are areas that cannot be seen during a visual inspection.Organic deposits may be present in these areas, which may impair function.Furthermore, we can determine visible deposits in the progav 2.0.The deposits did not affect the technical properties at the time of the examination.Deposits caused by natural substances found in the body, such as protein, blood, or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic deposits can impair the integrity of the valve.At the time of the examination, we were unable to determine how the aforementioned functional impairment occurred.Organic material in the patient's own cerebrospinal fluid may have temporarily affected the function and are known side effects of hydrocephalus therapy.Not every deposit in a shunt system leads to a deterioration in performance, depending on its location, quantity, and consistency.The examination of the returned product is complete with the preparation of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx643t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, the shunt system caused an underdrainage.The patient underwent a revision procedure on (b)(6) 2025.No patient complications were reported as a result of the revision procedure.The complained product was sent to the manufacturer for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25
Common Device Name
4041906502010
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key22554383
Report Number3004721439-2025-00207
Device Sequence Number18588152
Product Code JXG
UDI-Device Identifier04041906502010
UDI-Public4041906502010
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX643T
Device Catalogue NumberFX643T
Device Lot Number20081685
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2025
Initial Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexMale
Patient Weight33 KG
-
-