ARTHREX, INC. BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM; BURR,SURGICAL Back to Search Results
Model Number BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(4) 2025, a sales representative via phone that (2) ar-300-b201 burrs broke.This occurred during a cheilectomy procedure on (b)(6) 2025 while shaving the first burr snapped, the fragment was removed and the second one was attempted which also broke.The fragments were retrieved and the case proceeded using an ar-300-b202.This resulted in a 12-15 minute delay in which it is unsure if additional anesthesia was administered.There was no harm on the patient.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint allegation is confirmed based on the customer-provided picture.Visual evaluation of the customer provided photo noted the device was broken.Refer to attachments.Based on the information provided, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to applying excessive mechanical forces and/or prying/leveraging the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Common Device Name
BURR,SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22554773
Report Number1220246-2025-02978
Device Sequence Number15847056
Product Code LXH
UDI-Device Identifier00888867028654
UDI-Public(01)00888867028654(11)250201(17)300131(10)085048
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Device Catalogue NumberAR-300-B201
Device Lot Number085048
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date06/26/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2025
Supplement Date Received by Manufacturer10/16/2025
Initial Report FDA Received Date07/21/2025
Supplement Report FDA Received Date10/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-