Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during insertion and/or improper bone prep.The bone quality of the patient and the information of the instrument used to prepare the bone socket for insertion of the implant were not provided.The dfu for this device, dfu-0087-eo revision 4, gives the following warning: 6.Bioabsorbable only: attempting implantation into hard, dense bone and/or drilling/punching smaller diameter holes than recommended may cause failure (breakage) of the implant during insertion.
|