ETHICON INC. EB EXC GRN 30IN 3-0 S/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Catalog Number X832H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an urology procedure (b)(6) 2025 and suture was used.During the procedure, i am writing this email to give some feedback shared by our urology surgeons regarding the ethibond 3-0.We have done penile plication procedures recently, and for the past two occasions, the 3-0 ethibond rb-1 has been breaking when they are tying a knot.While a 3-0 ethibond is quite a small suture, the surgeons believe it should not be breakable by their hands when tying a knot.We opened two sutures today, the 3-0 ethibond on rb1 ref no: (b)(4), lot: sjbdsz, and 3-0 ethibond sh ref no.(b)(4), lot: tlbeqr, and we were able to break it with medium effort.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - device evaluation pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Additional information was requested, and the following was obtained via: - have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).None known.- how many devices demonstrated the alleged deficiency in each case? please specify by product code and lot number: unknown.- were there patient consequences? if yes, please explain.No.As of today, samples have been recovered for testing from manukau surgical centre.¿ product code x872h- lot sjbdsz: (b)(4).¿ product code x832h- lot tlbeqr: (b)(4).Additional information: was surgery delayed due to the reported event? unknown, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, (b)(6).Device property of: none, device in possession of: none, (b)(6).Device property of: none, device in possession of: none.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.One open box with twenty-two representative unopened samples were received for analysis.Product code x832h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.Following the sampling plan, a visual inspection was conducted on thirteen samples.The packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.A functional test was performed using instron equipment and tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigation summary: the product was returned to ethicon for evaluation.One open box with seven representative unopened samples was received for analysis.Product code x872h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB EXC GRN 30IN 3-0 S/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22555692
Report Number2210968-2025-08416
Device Sequence Number9700175
Product Code GAT
UDI-Device Identifier10705031058187
UDI-Public(01)10705031058187(17)280930(10)TLBEQR
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberX832H
Device Lot NumberTLBEQR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date07/21/2025
Supplement Report FDA Received Date08/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-