Investigation of this event is pending and a supplemental report will be sent upon its completion.Concomitant medical products: dvab1d1 icd implanted: (b)(6) 2016, 1103 vad implanted: (b)(6) 2020.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
A supplemental report is being submitted for device evaluation.Product event summary: the battery (b)(6) was not returned for evaluation.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Log file analysis revealed that the controller in use during the reported event (b)(6) contained a feature that records whether a power source experienced a communication error or a disconnection within each 15 minute interval.Analysis of the data log file revealed two instances involving (b)(6) where the battery's relative state of charge (rsoc) value was logged between 101-201, logged on (b)(6) 2025 at 00:45:20 and 11:46:42, which is indicative of a communication error.A communication error will trigger a power disconnect alarm if the other power source is a power adapter or a battery with an rsoc greater than 25%.However, review of the alarm file revealed no power disconnect alarms were logged within the analyze period.Additionally, analysis of the data log file revealed several premature power switching events that were due to momentary disconnections involving (b)(6) and additional batteries.As a result, the reported event was confirmed.The associated batteries had been lubricated prior to release.Possible root causes of the communication errors, and resulting power disconnect alarms, can be attributed to momentary disconnections on the communication pins of the controller, the controller not receiving responses from the batteries, and/or due to the packet error checking method detecting bit errors.Bas ed on an investigation conducted under capa pr00574181, a possible root cause of the reported premature power switching event can be attributed to momentary disconnections due to fretting corrosion of the controller-port/power-source pins, contamination on the controller receptacle sockets/power source pins, and/or a large spring gap between the controller receptacle spring and trough caused by a damaged receptacle.Even though this capa is now closed, the associated batteries falls in scope of this capa.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|