This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h4, h6 investigation summary according to the information received, "it was reported that the hospital was complaining about trays due to chipping, peeling, cracking and outer rubber peeling off.They were asking for replacements for sterilization requirements.The event did not happen in surgery." the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned sample revealed that the plastic coating on the brackets of the attune patella prep was cracked and peeling off.The overall appearance of the device exhibits evidence of constant use, servicing, and repetitive sterilization cycles.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune patella prep would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device due to the wear and tear condition.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|