• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS SMARTSITE PUMP INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS SMARTSITE PUMP INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 10010454
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd alaris smartsite pump infusion set was leaking the following information was received by the initial reporter with the following verbatim it was reported by customer that they noticed micron filter was wet, cleaned filter.Micron filter noted to be leaking.
 
Additional Manufacturer Narrative
The customer reported that tpn filter was leaking, and returned one used set of material 10010454, lot unknown.There is also a competitor trifuse set included as well.Upon initial visual examination, the sets had no defects or abnormalities.The bd set was infused with water, to flush the system.The supplier of the filter component suggests a bubble test, which consists of infusing air into the set while submerged underwater.This test is for the air vent membrane of the filter, which is the main culprit involved in the filter leaks.The bubble test showed that the filter was working properly, which shows the filter's air vent membrane hydrophobicity was compromised.Another round of flushing plus dry time was able to restore the vents to their hydrophobic qualities.This means the filter did not leak after the cleaning process, and it rules out any manufacturing error with the filter.The root cause could not be determined.Potential scenarios exist where the end user may infuse drugs/solutions into the filter that can weaken or compromise the hydrophobic properties of the air vent, such as low tension fluids like alcohol and lipids.A device history record review could not be performed because the lot number is unknown.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE PUMP INFUSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22560170
Report Number9616066-2025-02061
Device Sequence Number9473925
Product Code FPA
UDI-Device Identifier07613203015806
UDI-Public(01)07613203015806
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number10010454
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date07/21/2025
Supplement Report FDA Received Date07/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-