| Catalog Number |
762165 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using paxgene® blood rna tube, the solution within the tube backflowed into the patient.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 1 photo for investigation.The photo was reviewed and the indicated failure mode for backflow with the incident lot was not observed.Additionally, 10 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to backflow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode backflow.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
|
| |
|
Event or Problem Description
|
|
It was reported while using paxgene® blood rna tube, the solution within the tube backflowed into the patient.No adverse impact was reported regarding the patient or user.
|
| |
|