MAUDE Adverse Event Report: ABBOTT MEDICAL EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Pain Relief (2388)
Date of Event 07/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
It was reported that the patients ipg became unable to communicate with external devices.As a result, surgical intervention may be undertaken to address the issue.
 
Additional Manufacturer Narrative
An inoperable implant due to not charging the device was reported to abbott.The device has not been returned for analysis.A review of documentation supplied with the implant states that the implant must be charged every 30 days to avoid battery depletion.Based on the information received, the cause of the reported incident is consistent with user error.
 
Event or Problem Description
Additional information indicates that surgical intervention was undertaken on (b)(6) 2025 wherein ipg was explanted and replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Common Device Name
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key22561736
Report Number1627487-2025-03619
Device Sequence Number17119619
Product Code LGW
UDI-Device Identifier05415067019554
UDI-Public(01)05415067019554(10)5304880(17)160226
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/26/2016
Device Model Number3788
Device Catalogue Number3788
Device Lot Number5304880
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date07/21/2025
Supplement Report FDA Received Date08/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR (X2); SCS LEAD (X2)
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight95 KG
-
-