| Model Number |
CL-945 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during suturing, the needle detached from the suture and remained buried in the uterus.The needle was retrieved by searching manually by touch. an additional new suture was used without issue.The event resulted in extended surgical time.There was no patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during cesarean section under general anesthesia, the needle detached from the suture and remained buried in the uterus.The needle was retrieved by searching manually by touch.An additional new suture was used without issue.The event resulted in extended surgical time of 20 minutes.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a2, a4, b5, d9, g3, h3, h6.H3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the needle was broken at the swage and the suture was still attached to the broken swage.Upon microscopic inspection, instrument marks were observed near the break site.It was reported that the needle detached from the suture and remained buried in the uterus.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the customer is advised to grasp the needle by the middle section, where the diameter is greatest.Grasping the needle near the swaged end or the tip could cause bends, breaks, or damage the connection between the needle and suture.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|