| Catalog Number |
151620705 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the bearing would not seat, the item was wasted.Affected side: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the bearing would not seat, the item was wasted.The product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photographs attached were reviewed, however they do not represent the reported complaint.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.A manufacturing record evaluation was performed for the finished device product description: attune cr fb insrt sz 7 5mm product code: 151620705 lot number: d25040439 and no non-conformances or manufacturing irregularities were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product description: attune cr fb insrt sz 7 5mm product code: 151620705 lot number: d25040439 and no non-conformances or manufacturing irregularities were identified.Device history review: a manufacturing record evaluation was performed for the finished device product description: attune cr fb insrt sz 7 5mm product code: 151620705 lot number: d25040439 and no non-conformances or manufacturing irregularities were identified.H11 additional narrative: added: b5.
|
| |
|
Event or Problem Description
|
|
Additional information received states that the affected side involved in this event was the right knee.
|
| |
|
Search Alerts/Recalls
|