• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH RIA 2 BONE HARVESTING KIT 520MM STERILE; REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RIA 2 BONE HARVESTING KIT 520MM STERILE; REAMER Back to Search Results
Catalog Number 03.404.000S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure the extraction kit of the ria system is passed, the proper connections were made to pass the reamer and at the time of rhyming the medullary channel it was noticed that the yellow seal, the one that is responsible for not leaving the water, was melting.The specialist was informed to stop so it could be changed.It was then noticed that the blue plastic was also melting.For this novelty there was no discomfort on the part of the doctor.The doctor reports that he knew that the ria system was not indicated for this procedure but that he needed it to collect the pathology.The surgery was completed successfully, without affecting the patient and without alterations in the surgical times.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product was not returned to medtech orthopedics; however, photos were provided for review.The photo investigation revealed that ¿03.404.000s ria 2 bone harvesting kit 520mm sterile¿ has broken.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the ria 2 bone harvesting kit 520mm sterile would contribute to the complained device issue.Based on the investigation findings, ria 2 bone harvesting kit 520mm sterile was broken because of component failure, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review : part # 03.404.000s.Lot # 37840p6.Manufacturing site: werk selzach logistik.Release to warehouse date: 03.Oct.2024.Supplier: (b)(4).Expiration date: 01.Oct.2026.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIA 2 BONE HARVESTING KIT 520MM STERILE
Common Device Name
REAMER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22563741
Report Number8030965-2025-07324
Device Sequence Number3994271
Product Code HTO
UDI-Device Identifier10886982274236
UDI-Public(01)10886982274236(17)261001(10)37840P6
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K993335
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2026
Device Catalogue Number03.404.000S
Device Lot Number37840P6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/28/2025
Supplement Date Received by Manufacturer07/25/2025
Initial Report FDA Received Date07/22/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-