Model Number IT1032800B
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
The device was evaluated in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There is no allegation of serious or permanent harm or injury.The device is not reported to be in patient use.During the evaluation of the device, at a third-party service center the device was visually inspected and found evidence of foam degradation.During the evaluation, the type of contamination has been documented as dust on the blower box.Additionally, an error indicating the detachable battery is connected but the device does not recognize the connection was observed.The technician recommended replacing the device's detachable battery to address the issue.The device's missing rubber feet and o-rings require replacement.No further investigation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22565423
Report Number2518422-2025-108056
Device Sequence Number13066356
Product Code CBK
UDI-Device Identifier00606959042926
UDI-Public(01)00606959042926
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIT1032800B
Device Catalogue NumberIT1032800B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2025
Initial Report FDA Received Date07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/09/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1956-2021
Patient Sequence Number1
Patient SexUnknown
-
-