| Catalog Number |
123518A |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the patient noticed a leak in the foley catheter balloon on the way home after catheterization.Upon examination, the doctor confirmed that the balloon was indeed damaged.The defective product was not retained, and a new set was used to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive as no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.No sample was returned for evaluation.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).The labeling and instructions for use were reviewed and found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that, the patient noticed a leak in the foley catheter balloon on the way home after catheterization.Upon examination, the doctor confirmed that the balloon was indeed damaged.The defective product was not retained, and a new set was used to resolve the issue.
|
| |
|