Additional information was provided in d.9., h.3., h.6.And h.11.The device with the lens was returned loose in the carton.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Inadequate viscoelastic was observed in the device.The plunger was advanced into the nozzle area and has overrode the lens.The lens was slightly advanced inside the lens loading area.The leading haptic was extended along the right side of the plunger.The trailing haptic was folded onto the optic surface along the left side of the plunger.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.A qualified viscoelastic was indicated.A plunger underride was observed.The lens was still within the loading area.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).An inadequate amount of viscoelastic was observed in the device.The ifu instructs: fully insert the ophthalmic viscosurgical device (ovd) cannula, perpendicular to the device, through the viscoelastic port, located in the lens stop portion of the device.Fill the device until ovd can be observed flowing to the ¿fill-to¿ line on the nozzle tip.This will require approximately 0.2 millimetre of ovd.Only use an company ovd qualified for use with the company pre-loaded delivery system that has been allowed to come to the operating room temperature.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device.Plunger override may occur: due to rapid advancement faster than the ifu recommended rate.If inadequate viscoelastic is placed in the device, this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to override the lens.If the operating room temperature is too high (greater than 23 degrees celsius or 73 degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
|