| Model Number |
VICM5 13.2(-9.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no additional similar complaint type events within associated lots were found.Iop elevation from baseline and secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault and elevated iop without pupil block and angle closure.Due to excessive vault and elevated iop without pupil block and angle closure, the lens was exchanged for a different model but shorter length lens.The problem was resolved.Cause of the event is unknown.
|
| |
|
Search Alerts/Recalls
|