| Catalog Number |
365985 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported prior to use of bd microtainer® pst¿ tubes with lh (lithium heparin), an unspecified number of tubes had the gel located at the top of the tube or in the cap.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received two (2) photos of the customer samples for investigation.The photos were visually evaluated and gel smearing was confirmed; the barrier gel was smeared on the wall of tube.Additionally, fifty (50) retention samples were visually inspected for gel defects and no issues were observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for gel smearing based on the photos of the samples provided.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported prior to use of bd microtainer® pst¿ tubes with lh (lithium heparin), an unspecified number of tubes had the gel located at the top of the tube or in the cap.There was no health impact or consequences reported.
|
| |
|