Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A review of quarterly events indicated that for twenty-one (21) test events using the echo lumena automated blood bank system, a malfunction occurred resulting in the instrument producing incorrect results.
 
Additional Manufacturer Narrative
No common specific root cause or defect was identified.The conclusion(s) code(s) reported herein are assigned according to the facts presented by the affected customer and immucor's assessment and investigation of those facts.When possible, immucor attempts to obtain the actual samples and reagents involved; retention reagents of the same lot that was involved in the event may be used during the investigation if indicated.Also, when possible, immucor uses remote access to review the relevant data archived on the instrument.The events reported on this quarterly malfunction summary report are limited to those in which no patient harm occurred, and no design defect (or other systemic problem) was identified.There is no specific action that users are expected to take to mitigate the reported device failure/malfunction other than to ensure that preventative maintenance is performed as indicated in the instrument instructions.Immucor will continue to track and trend performance and operational issues such as described in this quarterly malfunction summary report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO LUMENA
Common Device Name
AUTOMATED BLOOD BANK SYSTEM
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30071
Manufacturer (Section G)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30071
MDR Report Key22573448
Report Number1034569-2025-00017
Device Sequence Number9784602
Product Code KSZ
UDI-Device Identifier10888234002338
UDI-Public10888234002338
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
BK070016
Exemption NumberVMSR
Number of Events Summarized21
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Periodic
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-