Batch review performed on 16 july 2025 lot.2305365: (b)(4) items manufactured and released on 03-july-2023.Expiration date: 17-june-2028.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with a similar reported event during the period of review.Additional component invovled and revised: reverse shoulder system 04.01.0212 lat.Glenosphere 42xø27 (k193175) lot.2345639 : (b)(4) items manufactured and released on 07-march-2024.Expiration date: 21-febr-2029.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Based on the information available no definitive root cause can be established, while there is no indication that any potential issue with the device may have caused or contributed to the event, and the device history record review does not indicate any potential manufacturing related issue.
|
|
Primary surgery performed on (b)(6) 2025.1st revision surgery performed on (b)(6) 2025 due to joint luxation; liner and glenosphere revised.2nd revision surgery performed on (b)(6) 2025 due to joint luxation; liner and glenosphere revised.3rd revision surgery performed on (b)(6) 2025 due to joint luxation; liner, glenosphere, and metaphysis revised.Presently, on (b)(6) 2025, the 4th revision surgery has been performed due to joint luxation; patient converted to hemiarthroplasty - glenosphere and baseplate removed, liner and metaphysis replaced with anatomical components.Patient history: primary surgery performed on (b)(6) 2025.1st revision surgery performed on (b)(6) 2025 due to joint luxation; liner and glenosphere revised.(mdr 2025-00721).2nd revision surgery performed on (b)(6) 2025 due to joint luxation; liner and glenosphere revised.(mdr 2025-00720).3rd revision surgery performed on (b)(6) 2025 due to joint luxation; liner, glenosphere, and metaphysis revised.(mdr 2025-00723).4th revision surgery performed on (b)(6) 2025 due joint luxation; patient converted to hemiarthroplasty - glenosphere and baseplate removed, liner and metaphysis replaced with anatomical components (mdr 2025-00719).5th revision surgery performed on (b)(6) 2025 due to intra-prosthetic dislocation (metal humeral head and double eccenter) - humeral head and the double eccentric revised and surrounding bone trimmed using a rongeur.(mdr 2025-00663).
|