• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SHERIDAN/CF ET TUBE,CF,8.0; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SHERIDAN/CF ET TUBE,CF,8.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Catalog Number 5-10116
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "cuff would not stay inflated after intubation.Each cuff was inflated prior to intubation to test.Associated complaints 3003898360-2025-00597 and 3003898360-2025-00598.
 
Additional Manufacturer Narrative
(b)(4) the customer returned one unit 5-10116 et tube, cf,8.0 for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned device revealed that the sample appeared typical.No defects or anomalies were observed.Functional inspection was performed on the returned device to test for proper inflation and deflation.A lab inventory syringe was filled with air and attached to the pilot balloon.Upon injection of air, the balloon cuff was unable to stay inflated.The device was submerged underwater in order to test for leaks.Upon inflation, the device leaked from two small holes in the balloon cuff.The reported complaint of "leak - device leak - cuff" was confirmed based upon the sample received.The returned et tube was found to have two holes in the balloon cuff which prevented it from being able to stay inflated.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHERIDAN/CF ET TUBE,CF,8.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
elaine cully
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key22574183
Report Number3003898360-2025-00571
Device Sequence Number21305771
Product Code BTR
UDI-Device Identifier04026704396825
UDI-Public(01)04026704396825(11)231012(17)281010(10)73K2300382(21)
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5-10116
Device Lot Number73K2300382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2025
Initial Report FDA Received Date07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED
Patient SexUnknown
-
-