• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 367287
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.The manufacturing location for this product is nipro.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.
 
Event or Problem Description
Report 1 of 2 it was reported that while using one bd vacutainer® safety-lok¿ blood collection set, blood leakage occurred at the end of non-patient needle causing contamination on the bed sheets, patients' clothes and nurses' clothes.There was no health impact or consequence reported.
 
Additional Manufacturer Narrative
Investigation summary : bd had not received samples or photos of the defects for investigation.Therefore, 50 retention samples from bd inventory were evaluated by visual examination and nothing abnormal was found from the rubber sleeves.Also, functional testing was conducted on 8 retained samples using bd tubes and nothing abnormal was observed as no rubber sleeve retracting defects or side-piercing defects occurred during the test, all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of sleeve leakage based on retains testing.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
Report 1 of 2.It was reported that while using one bd vacutainer® safety-lok¿ blood collection set, blood leakage occurred at the end of non-patient needle causing contamination on the bed sheets, patients' clothes and nurses' clothes.There was no health impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key22576008
Report Number2243072-2025-00907
Device Sequence Number21222388
Product Code JKA
UDI-Device Identifier30382903672876
UDI-Public(01)30382903672876(17)260630(10)24G03A2
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K980414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue Number367287
Device Lot Number24G03A2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date07/23/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-