• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION PROCISE EZ VIEW COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION PROCISE EZ VIEW COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number EIC8875-01
Medical Device Problem Codes Break (1069); Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during an ent procedure, the metal wire of the procise view coblator wand broke.The procedure was completed using a s+n back up device.There was a delay of less than 30 minutes.No further complications were reported.
 
Additional Manufacturer Narrative
H11: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.However an analysis of the customer provided image found a wand being held with the electrodes facing upwards, the bottom electrode is missing most of its body at the center, a part of the stand can be appreciated.The tip shows signs of wear by use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (1) using the wand above default output settings, thus causing the electrodes to become eroded extensively.2) pressing the tip against hard or bony surfaces).Based on this investigation, the need for corrective action is not indicated.
 
Additional Manufacturer Narrative
H11, h3, h6: the reported device was received for evaluation.A visual inspection observed the returned instrument shows no manufacturing abnormalities.Product was out of the original packaging.No packaging returned.The tip is worn from use.The electrodes are thin from use.The bottom electrode is missing most of its body from use.A functional evaluation of the device found that the opened device was tested and plugged into the controller and registered settings (7,3).The wand can produce plasma with its remaining electrode and had no issues activating coag.An analysis of the customer provided image found a wand being held with the electrodes facing upwards, the bottom electrode is missing most of its body at the center, a part of the stand can be appreciated.The tip shows signs of wear by use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (1) hitting the device tip against a hard surface.2) using the device as a lever to enlarge surgical site.3) mechanical displacement of tissue through applied force.4) activating the device above recommended settings for prolonged periods of time).Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCISE EZ VIEW COBLATOR II
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key22576044
Report Number3006524618-2025-00385
Device Sequence Number18567000
Product Code GEI
UDI-Device Identifier00817470000313
UDI-Public00817470000313
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070374
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEIC8875-01
Device Lot Number2161664
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer09/02/2025
11/21/2025
Initial Report FDA Received Date07/23/2025
Supplement Report FDA Received Date09/03/2025
11/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-