| Catalog Number |
367921 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6) hospital.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® sst¿ blood collection tube, the tube underfilled.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received one photo for investigation, which was visually evaluated and confirmed the customer's indicated failure mode.Additionally, a total of 100 retained samples were inspected, and none showed visible defects.Furthermore, a functional test was performed on 10 retained samples, and all tubes were found to be within specification limits.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: draw volume.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using one bd vacutainer® sst¿ blood collection tube, the tube underfilled.There was no health impact or consequence reported.
|
| |
|