Product analysis: as found condition: the axium prime device was returned for analysis within a shipping box and within a sealed, labeled, plastic biohazard pouch.The unknown micro catheter used in the event was not returned for analysis.Visual inspection/damage location details: the pusher was found broken at the break indicator with the release wire not retracted.The axium prime implant coil was found stretched and damaged with the polypropylene filament broken.Testing/analysis: the exposed release wire was retracted, detaching the implant coil medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received information regarding issues with two different axium coils which occurred during the same case.The patient was undergoing a coil embolization procedure to treat an internal carotid artery unruptured saccular aneurysm.The aneurysm max diameter was 15mm and the neck diameter was 7mm.Blood flow and vessel tortuosity were normal.It was reported that the devices were prepared as indicated in the instructions for use (ifu).During the procedure the tail of the apb-1-2-3d-es coil was stretched.The coil was removed and replaced to continue the procedure.During the same case, another axium coil (qc-10-30-3d) could not be detached.The coil was again replaced to complete the procedure successfully.There was no harm or injury to the patient.Additional information received reported there were no issues that resulted in the coil stretch to the apb-1-2-3d-es.Regarding the coil qc-10-30-3d, no instat detacher was used.No less than 3 manual attempts to detach.
|