This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: part: 03.037.018.Lot: 9837968.Manufacturing site: werk hagendorf.Release to warehouse date: 29 february 2016.Supplier: n/a.Expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found both of the color markers were missing.Potential cause for this condition can be traced to maintenance and/or sterilization procedures, however, with available information the complete potential cause can not be determined.According to the document cleaning and sterilization instructions, the following precautions should be taken: all devices must be thoroughly cleaned and inspected prior to sterilization.Long, narrow lumens, blind holes, moving and intricate parts require particular attention during cleaning and inspection.During cleaning, only use cleaning agents that are labelled for use on medical devices and in accordance with the manufacturer¿s instructions (e.G.Temperature, contact time, and rinse time).Cleaning agents with a used dilution ph of within 7¿9.5 are recommended.Highly alkaline conditions (ph >11) can damage components/devices, such as aluminum materials.Do not use saline, environmental disinfection (including chlorine solutions) or surgical antiseptics (such as iodine- or chlorhexidine-containing products).Do not use a cleaning aid that can damage the surface of instruments such as steel wool, abrasive cleaners or wire brushes.Additionally, the prong component of the shaft was slightly bent to the outside.This observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional evaluation was performed with mating device 03.037.017 blade/screw guide sleeve and assembly showed resistance due to bent condition.The overall complaint was confirmed as the observed condition of the 3.2mm wire guide sleeve would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The drawings reflecting the current and manufactured revisions were reviewed.
|