|
Manufacturer ref# (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
A healthcare professional reported that during an endovascular embolization, an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9236682) was impeded in the proximal end of a prowler select plus 150/5cm microcatheter (606s255x, lot unknown) and could not advance any more.The doctor removed the stent and microcatheter from the patient and switched new devices to complete the procedure.Additional event information received on 23-jul-2025 indicated that no torque device was used.There was no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.There was no procedure prolongation/delay due to the event.
|
|
Manufacturer ref# (b)(4) updated sections on this medwatch: b4, g3, g6, h2, h3, h6 and h11.Complaint conclusion: a healthcare professional reported that during an endovascular embolization, an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9236682) was impeded in the proximal end of a prowler select plus 150/5cm microcatheter (606s255x, lot unknown) and could not advance any more.The doctor removed the stent and microcatheter from the patient and switched new devices to complete the procedure.Additional event information received on 23-jul-2025 indicated that no torque device was used.There was no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.There was no procedure prolongation/delay due to the event.The device was returned to j&j medtech for further evaluation.A non-sterile enterprise2 4mmx23mm was received contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and it was noted that the stent was already detached from the unit.The delivery wire and the introducer were in good condition (i.E., no kinks, bents, or elongations).The stent was inspected under a microscope, and no damage was found (i.E., no kinks, no bents, or broken struts).Both distal ends can be noted completely flared.The issue regarding a stent being impeded at proximal end of the microcatheter cannot be evaluated through functional testing.The stent must be inside the introducer tube to perform the functional analysis.Additionally, none of the returned components present damages that suggest that they were forcibly advanced.With the limited information available, a conclusive cause cannot be determined; however, it is possible that clinical and procedural factors, including device manipulation and operator's technique, may have contributed to the reported failure.At this time, there is no evidence to support that the issue reported in the complaint is a result of a defect inherently related to the device.The stent detachment was not originally reported in the complaint, therefore, the exact time when this condition occurred cannot be determined.This condition is not considered related to the reported issue.Lake region medical did review the device history records relative to the manufacturing, inspecting and packaging of the lot 9236682.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations: ¿ if resistance is met during manipulation, determine the cause of resistance before proceeding.Do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance a new one.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|