• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA GLIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA GLIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number GS3506
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility reported that the procedure was complicated by a fractured guidewire with the guidewire fragment lodged in a subintimal location along the left iliac artery.The attempts to snare and remove the fragment were not successful.Eventually, they decided that the safest thing for the patient was to leave the wire fragment in place.The fragment would have been secured by its subintimal location.The wire fragment was further secured with the deployment of adjacent balloon-expandable stents.The procedure was arterial angioplasty and atherectomy.No blood loss was reported.The reported event did not result in patient injury and/or require medical or surgical intervention.Additional information was received on 14jul2025: the size of fragment left in patient was 13 cm.Within a subintimal tract with tortuosity and resistance with torque applied to the wire using a torque device.The wire had been advanced through 5f kumpe and omni select catheters.The patient was in stable condition.There were no plans to remove wire fragments since they placed stents to secure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya, shizuok 418-0 015
JA  418-0015
MDR Report Key22578816
Report Number2243441-2025-00039
Device Sequence Number9237295
Product Code DQX
UDI-Device Identifier04987350773593
UDI-Public(01)04987350773593(17)260930(10)241024
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/23/2025
Report Date (Section F) 07/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberGS3506
Device Catalogue NumberRF*PA35153A
Device Lot Number241024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/02/2025
Device Age8 MO
Event Location Hospital
Date Report to Manufacturer07/04/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/23/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
Date Report Sent to FDA07/23/2025
-
-