| Catalog Number |
3009PX-ZMX-600 |
| Medical Device Problem Code |
Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report summarizes 7 malfunction events(s), where it was reported the devices experienced zoom suddenly stops or zoom unexpectedly retracts/disengages during use.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.7 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Additional Manufacturer Narrative
|
|
In accordance with the ¿final guidance on medical device reporting for manufacturers¿ issued on november 7, 2016, stryker medical will no longer report the hazard of zoom suddenly stops for our bed lines (product code fnl), as this hazard has not caused or contributed to any serious injuries.While no new mdrs will be generated for this hazard moving forward, additional supplemental records may be submitted for past reported events if new information becomes available.Should a new serious adverse event occur attributed to this hazard, mdr reporting will resume.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|