Model Number EE2110X15B
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo ventilator was returned to a third-party service center for service.There was no harm or injury reported.A ventilator service required alarm was found in the device error log indicating the internal or detachable li-ion battery was not charging due to a hardware fault for greater than, or equal to 1 minute.The unit was tested with a gold detachable battery because the unit came without one.The device was left running for 4 hours and no alarm appeared.It is possible that the detachable battery used by the customer is faulty and needs to be replaced.This issue is reportable because the system cannot charge the internal battery under this condition.This device has been serviced, inspected, tested, met acceptance criteria in accordance with all of the applicable customer requirements as defined in the contractual agreement with the third-party servicer.The device history record has been reviewed and approved by the third-party servicer's quality assurance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22579178
Report Number2518422-2025-108164
Device Sequence Number18450940
Product Code CBK
UDI-Device Identifier00606959055544
UDI-Public(01)00606959055544
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEE2110X15B
Device Catalogue NumberEE2110X15B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2025
Initial Report FDA Received Date07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-