| Model Number |
620-2221411-01 |
| Medical Device Problem Codes |
Failure of Device to Self-Test (2937); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including stressing with numerous power cycles without duplicating the customer's report.The accessories were not returned foe evaluation.The processor/bridge/pace board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for defib function.The user was subsequently unable to select the energy level due to unresponsive keypad response.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|