Model Number 102996
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Migration (4003)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 07/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A patient implanted with an evoke spinal cord stimulation (scs) system was evaluated for inadequate pain relief.The patient reported increasing stimulation in the abdomen area and one (1) event of unexpected interruption of therapy in closed loop.X-ray imaging was obtained which confirmed a lead migration.The patient was reprogrammed to temporarily restore therapy, pending a lead revision to resolve the lead migration.A lead revision procedure was completed therapy restored to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOKE ACTIVE ANCHOR KIT (US)
Common Device Name
SCS ANCHOR
Manufacturer (Section D)
SALUDA MEDICAL PTY LTD
5 eden park drive
macquarie park, nsw 2113
AU  2113
MDR Report Key22579632
Report Number3016111136-2025-00117
Device Sequence Number7091626
Product Code LGW
UDI-Device Identifier09352307001718
UDI-Public09352307001718
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/23/2025
Report Date (Section F) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/06/2026
Device Model Number102996
Device Catalogue Number3043
Device Lot Number9016910305
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/07/2025
Event Location Hospital
Date Report to Manufacturer07/07/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/23/2025
Date Device Manufactured02/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-