• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION THORATEC® HEARTMATE® XVE AND HEARTMATE II®, APICAL CORING KNIFE; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION THORATEC® HEARTMATE® XVE AND HEARTMATE II®, APICAL CORING KNIFE; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1050
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event or Problem Description
It was reported that the surgeon preferred to use a stand along coring wife due to recent issues with coring knives.The knife only cut through the top layers of the cardiac muscle.Scissors were needed to remove all of the core from the ventricle.No harm reached the patient.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: evaluation of the returned coring knife, serial number (b)(6), could not confirm the reported cutting issue.A specific cause for the reported event could not be conclusively determined through this evaluation.The coring knife was returned in a plastic jar wrapped in surgical gloves.The handle and protective caps were not returned.Residue consistent with blood was present on the knife body and blade edge.Evidence of rust was observed on the blade edge, consistent with fluid contact during the implant surgery.The knife otherwise appeared unremarkable upon visual inspection.Microscopic inspection did not reveal any evidence of damage or defect.A cut test was performed with the returned coring knife and a polyurethane sheet.The knife was able to cut through this material without issue, comparable to a control knife.Review of the coring knife manufacturing documentation found no deviations from manufacturing or quality specifications.In addition, a corrective action/preventive action (capa) investigation was opened to further investigate issues related to the coring knife not being able to cut.The relevant sections of the device history records were reviewed and showed no deviations from manufacturing or quality assurance specifications.The current revision of the instructions for use (ifu) can be found on the eifu page of the abbott website.The heartmate 3 left ventricular assist system (lvas) ifu, is currently available.Chapter 1, "introduction", lists the apical coring knife as an optional component for device implantation.Chapter 5, ¿surgical procedures¿, provides information on preparing and handling the coring knife.This chapter also states that during the implant process, a complete backup system (implant kit and external components) must be available on-site and in close proximity for use in an emergency.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORATEC® HEARTMATE® XVE AND HEARTMATE II®, APICAL CORING KNIFE
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key22579633
Report Number2916596-2025-04770
Device Sequence Number12535532
Product Code DSQ
UDI-Device Identifier00813024010227
UDI-Public(01)00813024010227(17)260918
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/18/2026
Device Model Number1050
Device Catalogue Number1050
Device Lot Number9242645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2025
Supplement Date Received by Manufacturer02/19/2026
Initial Report FDA Received Date07/23/2025
Supplement Report FDA Received Date02/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
Patient Weight122 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-