| Model Number |
CR2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device¿s lid magnet was missing from the lid.As a result, the device may not automatically power on when the lid is opened.This may led to defibrillation therapy being delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer received a replacement lid and battery.The device was not returned to stryker for evaluation.The cause of the reported issue could not be determined.
|
| |
|
Search Alerts/Recalls
|