• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS PLASTIPAK LUER-LOK; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS PLASTIPAK LUER-LOK; PISTON SYRINGE Back to Search Results
Catalog Number 309658
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd syringe 3ml ll euro 200 s/c barrel cracked.The following information was provided by the initial reporter: this morning we observed two cracks in bd 3ml ll syringes: one caused the syringe to break while preparing a cytotoxic drug the other crack was noticed by a nurse after administering a cytotoxic drug sc --> a drop appeared outside the syringe.The batches that may be affected are: 5024932; 4352797; 4352796; 3073041.As the syringes contained cytotoxic products, we did not keep them.10-jul-2025 - additional information received reference number: 309658.
 
Additional Manufacturer Narrative
(b)(6) - supplemental mdr - barrel cracked.Since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASTIPAK LUER-LOK
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22583615
Report Number1213809-2025-00496
Device Sequence Number15773589
Product Code FMF
UDI-Device Identifier00382903096589
UDI-Public(01)00382903096589(11)241217(17)291130(10)4352796
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number309658
Device Lot Number4352796
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date07/23/2025
Supplement Report FDA Received Date08/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-