|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) bnst3211, (3i t3 with dcd non-platform switched tapered implant 3.25 x 11.5mm) for evaluation.Visual evaluation was performed, the implant had signs of use with bone debris on the external threads.The implant was fractured at the body.The bottom portion of the implant was not returned.Dhr review was completed for the subject lot number 2022101305.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022101305 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental : functional : fracture : implant¿.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors - occlusal loading.Therefore, based on the available information, a device malfunction did occur.The implant was fractured at the body.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|