• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAYENNE MEDICAL QUATTRO SUTURE PASSER NEEDLE; NEEDLE, SUTURING, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAYENNE MEDICAL QUATTRO SUTURE PASSER NEEDLE; NEEDLE, SUTURING, DISPOSABLE Back to Search Results
Catalog Number CM-9011
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, due to hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during surgery that while passing the needle through the cuff to the suture anchor, the tip broke off.Nothing could be located and there was no harm to the patient.Foreign body was not retained.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: h2; h3; h6; h10; h11.The reported event could not be confirmed due to lack of product/information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO SUTURE PASSER NEEDLE
Common Device Name
NEEDLE, SUTURING, DISPOSABLE
Manufacturer (Section D)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer (Section G)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
Manufacturer Contact
jennifer rapsavage
56 e. bell dr
warsaw, IN 46582
5745260384
MDR Report Key22586110
Report Number3006108336-2025-00028
Device Sequence Number9492924
Product Code GAB
UDI-Device Identifier00887868270147
UDI-Public(01)00887868270147(17)280319(10)67204503
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCM-9011
Device Lot Number67204503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2025
Supplement Date Received by Manufacturer12/18/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date12/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3006108336-2025-00027
Patient Sequence Number1
Patient SexUnknown
-
-