The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.There was no allegation reported by the patient.There was no report of serious patient harm or injury.There was no report of medical intervention.No additional information can be requested at this time.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found evidence of foam degradation.The device's downloaded logs were reviewed by the third-party service center.During evaluation dust/dirt contamination was observed.There were no errors code found.The third-party service center concludes that there was visible foam degradation.
|