• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP ELITE 14CM STRAIGHT ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP ELITE 14CM STRAIGHT ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number 5407120480
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 30 events were reported for this quarter.Product return status: 30 devices were evaluated in the field.Additional information: 30 devices were not labeled for single-use.30 devices were not reprocessed or reused.
 
Event or Problem Description
This report summarizes 30 events for the failure: flaked.30 events had problem identified during non-clinical procedure; there was no patient involvement.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale: 30 events were originally reported for this failure mode during the reporting quarter: however, 1 event was inadvertently excluded.31 reported events are included in this follow-up record.Product return status: 1 device was not available for evaluation.30 devices were evaluated in the field.Additional information: 31 devices were not labeled for single-use.31 devices were not reprocessed or reused.
 
Event or Problem Description
This report summarizes 31 events for the failure: flaked.31 events had problem identified during non-clinical procedure; there was no patient involvement.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Mfr report # 3015967359-2025-99014.Supplemental rationale, corrected data: b5, h6, h11.31 previously reported events were included in this follow-up record.Product return status.1 device was not available for evaluation.30 devices were evaluated in the field.Evaluation findings (results/components), 31 devices: material and/or chemical problem identified / coating material.Product disposition.1 device: product not received.30 devices: product disposition is not yet determined.
 
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between april 1 ¿ june 30, 2025.This report summarizes 31 events for the failure: flaked.- 31 events had problem identified during non-clinical procedure; there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELITE 14CM STRAIGHT ATTACHMENT
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key22586830
Report Number3015967359-2025-99014
Device Sequence Number18462226
Product Code HBE
UDI-Device Identifier07613327096583
UDI-Public07613327096583
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized31
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5407120480
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer06/30/2025
06/30/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date07/31/2025
10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-