• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP MD SERIES LONG STRAIGHT ATTACHMENT; UNIT, OPERATIVE DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP MD SERIES LONG STRAIGHT ATTACHMENT; UNIT, OPERATIVE DENTAL Back to Search Results
Catalog Number 5100015270
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes 3 events for the failure: detachment of device or device component.- 2 events had problem identified before clinical use/exposure; there was no patient involvement.- 1 event had problem identified during non-clinical procedure; there was no patient involvement.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events 3 events were reported for this quarter.Product return status 3 devices were received.Additional information 3 devices were not reprocessed or reused.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Mfr report # 3015967359-2025-99094.Supplemental rationale, corrected data: b5, h6, h11.3 previously reported events were included in this follow-up record.Product return status, 3 devices were received.Evaluation findings (results/components).2 devices: wear problem / housing.1 device: wear problem / housing, spring.Product disposition, 3 devices: scrapped by stryker.
 
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between april 1 ¿ june 30, 2025.This report summarizes 3 events for the failure: detachment of device or device component.- 2 events had problem identified before clinical use/exposure; there was no patient involvement.- 1 event had problem identified during non-clinical procedure; there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MD SERIES LONG STRAIGHT ATTACHMENT
Common Device Name
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key22587164
Report Number3015967359-2025-99094
Device Sequence Number18514727
Product Code EIA
UDI-Device Identifier04546540380289
UDI-Public04546540380289
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized3
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5100015270
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/30/2025
Supplement Date Received by Manufacturer06/30/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-