| Model Number |
620-2231511-01-66 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during initial testing; however, the device was put through extensive testing including bench testing and ecg stressing, without duplicating the customer's report.The customer's multi-function cable was not provided for the evaluation.It was recommended to the customer to discard the muflit-function cable used as a precaution.The defib receptable was replaced as a precaution and a new multifunction cable was provided to the customer.The device was recertified and returned to the customer.Analysis of the report of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "ecg monitoring failure" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Search Alerts/Recalls
|