HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1420
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that review of log file data revealed the controller exhibited a loss of power.It was also reported that it is unknown what caused the loss of power.The controller remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Continuation of d10: dtma1d1 crt-d implant date: (b)(6) 2021 5076-45 lead implant date: (b)(6) 2017 429688 lead implant date: (b)(6) 2017 694765 lead implant date: (b)(6) 2008.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for investigation summary.Product event summary: the controller ((b)(6)) was not returned for evaluation.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Log file analysis revealed one (1) controller power-up event, with an associated motor start event logged on (b)(6) 2025 at 12:37:13, indicating a loss of power to the controller.The data point logged prior to the loss of power revealed that (b)(6) was connected to power port one (1) with (b)(4) relative state of charge (rsoc) and that (b)(6) was connected to power port two (2) with (b)(4) rsoc.The data point recorded after the loss of power revealed that (b)(6) was connected to power port one (1) and (b)(6) was connected to power port two (2).The controller was without power for eighteen (18) seconds.No anomalies were recorded leading up to the loss of power.As a result, the reported event was confirmed.The most likely root cause of the losses of power to the controller can be attributed to a disconnection of both power sources and/or to an intermittent disconnection on one or both power sources.Capa pr00551638 investigated controller losses of power.Even though this capa is now closed, (b)(6) falls in scope of this capa.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22587743
Report Number3007042319-2025-01513
Device Sequence Number9392780
Product Code DSQ
UDI-Device Identifier00888707006859
UDI-Public(01)00888707006859(17)210630
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2021
Device Model Number1420
Device Catalogue Number1420
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2025
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date07/24/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/05/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD
Patient Age54 YR