| Catalog Number |
362760 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® cpt¿ cell preparation tube with sodium citrate, 5 samples exhibited poor separation after processing.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received one photo for investigation.The photo was reviewed and the indicated failure mode for poor barrier separation was observed.Bd was unable to determine the specific lot number associated with this complaint; therefore, a review of the device history record could not be conducted.Complaints for sample quality are under statistical control for the month of june 2025.At this time, further testing is not indicated.This complaint has been confirmed for the indicated failure mode poor barrier separation based on the photo provided.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® cpt¿ cell preparation tube with sodium citrate, 5 samples exhibited poor separation after processing.There was no health impact or consequences reported.
|
| |
|